Agilent Technologies 81101A Reference Manual page 72

50 mhz pulse generator
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

:VOLT[1]:HIGH
[:SOURce]:VOLTage[1][:LEVel][:IMMediate]:HIGH
Set & Query
Numeric
V with engineering prefixes.
Amplitude = High – Low
High – Low
Offset =
With low level, see
page 73
500 mV
–9.9 V to 10.0 V (50
Use this command to program the high level voltage of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of voltage, you first
have to execute the [:SOURce]:HOLD VOLTage command to enable the
[:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
• Specified current limits
• Actual OUTPUT impedance setting :OUTPut:IMPedance
• Actual expected load impedance setting
:OUTPut:IMPedance:EXTernal
To set the high level voltage:
:HOLD VOLT
:VOLT:HIGH 4.8V
72
2
into 50 )
Enable VOLTAGE subsystem
Set OUTPUT high level voltage to 4.8 V