Chapter 4 Features and Functions
Scanning
Automatic Channel Delays
If you do not specify a channel delay, the instrument selects a delay for
you. The delay is determined by function, range, integration time, and
AC filter setting as shown below.
DC Voltage, Thermocouple, DC Current (for all ranges):
Integration Time
Channel Delay
PLC >1
PLC
Resistance, RTD, Thermistor (2- and 4-wire):
Channel Delay
Range
(For PLC > 1)
100
2.0 ms
1 k
2.0 ms
10 k
2.0 ms
100 k
25 ms
1 M
30 ms
10 M
200 ms
100M
200 ms
AC Voltage, AC Current (for all ranges):
AC Filter
Channel Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
Frequency, Period:
AC Filter
Channel Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
Digital Input, Totalize
Channel Delay
0 sec
106
2.0 ms
1.0 ms
Range
100
1 k
10 k
100 k
1 M
10 M
100M
7.0 sec
1.0 sec
120 ms
0.6 sec
0.3 sec
0.1 sec
Channel Delay
For PLC
1.0 ms
1.0 ms
1.0 ms
20 ms
25 ms
200 ms