Agilent Technologies 34972A User Manual page 114

Data acquisition / switch unit
Table of Contents

Advertisement

Chapter 4 Features and Functions
Scanning with External Instruments
• Remote Interface Operation: The following program segment
configures the instrument for an externally controlled scan.
TRIG:SOUR TIMER
ROUT:CHAN:ADV:SOUR EXT
TRIG:TIMER 5
TRIG:COUNT 2
INIT
To configure the instrument for 4-wire external scanning, send the
following command.
ROUTe:CHANnel:FWIRe {OFF|ON}[,(@<ch_list>)]
114
Select the scan interval
Select the channel advance source
Set the scan interval to 5 seconds
Sweep the scan list 2 times
Initiate the scan