34905A/6A Dual 4-Channel Rf Multiplexers - Agilent Technologies 34972A User Manual

Data acquisition / switch unit
Table of Contents

Advertisement

34905A/6A Dual 4-Channel RF Multiplexers

These modules consist of two independent 4-to-1 multiplexers.The
channels in each bank are organized in a "tree" structure to provide high
isolation and low VSWR. Both banks have a common earth ground. This
module does not connect to the internal DMM. You can connect your
signals directly to the on-board SMB connectors or to the
SMB-to-BNC cables provided with the module.
98
Bank Switch
99
Bank Switch
NOTES:
The 34905A is used for 50 applications. The 34906A is used for 75 applications (mini
SMBs).
You can close only one channel per bank at a time on these modules; closing one channel
in a bank will open the previously closed channel. One channel in each bank is always
connected to COM.
This module responds only to the CLOSE command (OPEN does not apply). To OPEN a
channel, send the CLOSE command to another channel in the same bank.
Chapter 4 Features and Functions

34905A/6A Dual 4-Channel RF Multiplexers

11
12
Com1
13
14
21
22
Com2
23
24
4