Error Messages

Error Messages
Displaying the SCPI error queue
Error List
Error
0
100
101
102
103
104
105
106
107
108
109
110
111
112
113
156
Front Panel:
Press the Error key.
If errors appear, use the navigation
keys to scroll though the list.
The entire error queue is read, then emptied.
The following table documents the various error messages that the
power system supports:
Device-dependent Errors (these errors set Standard Event Status register bit #3)
No error
This is the response to the ERR? query when there are no errors.
Too many channels
You have specified more channels than are installed in the mainframe.
Calibration state is off
Calibration is not enabled. The instrument will not accept calibration commands.
Calibration password is incorrect
The calibration password is incorrect.
Calibration is inhibited by switch setting
Calibration mode is locked out by the calibration switch.
Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.
Unexpected output current
The measured output current is outside the acceptable range.
Zero measurement out of range error
The "zero" measurement value is outside the acceptable range.
Programming cal constants out of range
The programmed calibration constant is outside the acceptable range.
Measurement cal constants out of range
The measurement calibration constant is outside the acceptable range.
Over voltage cal constants out of range
The over voltage calibration constant is outside the acceptable range.
Wrong V+I
The instrument was unable to set the correct voltage or current value.
Aux vloc cal constants out of range
Calibration constants on the internal auxiliary local ADC are outside the acceptable range.
Aux vrem cal constants out of range
Calibration constants on the internal auxiliary remote ADC are outside the acceptable range.
Aux imon cal constants out of range
Calibration constants on the internal auxiliary imon ADC are outside the acceptable range.
SCPI Command:
SYST:ERR?