Series 4060
Threshold Alarm 2:
 Can be configured as high (actuates when
concentration is above threshold), or low (actuates
when concentration is below threshold).
 Can be configured as failsafe or non-failsafe
 Can be configured as non-latching
 Can be configured out (defeated).
System Alarm:
 Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be
defeated.
Note: Reset by pressing the STANDBY button to remove power.
Then press STANDBY again and any other button except
SYSTEM to resume. Further detail can be found in
Chapter 4, Section 4.6.19.

Table 3-4: Alarm Relay Contact Pins

Pin
Contact
45
Threshold Alarm 1, normally closed contact
28
Threshold Alarm 1, moving contact
46
Threshold Alarm 1, normally open contact
42
Threshold Alarm 2, normally closed contact
44
Threshold Alarm 2, moving contact
43
Threshold Alarm 2, normally open contact
36
System Alarm, normally closed contact
20
System Alarm, moving contact
37
System Alarm, normally open contact
Teledyne Analytical Instruments
Installation
33