Operation
GROUP SETUP
timing and measurement ranges. Do not change these.
Used to establish password protection or change the
PSWD:
existing password.
LOGOUT:
the analyzer settings.
STANDBY
maintains power to internal circuitry.
Any function can be selected at anytime. Just scroll through the
MAIN MENU with the DOWN/UP keys to the appropriate function, and
ENTER it. The analyzer will immediately start that function, unless
password restrictions have been assigned. (Password assignment is
explained in Section 4.6.17).

4.6.5 Span

The SPAN function is used to calibrate the analyzer. Span can be
performed either manually or automatically with the autocal function if
equipped.
The analyzer is calibrated using span gas as described in Section 4.1.
This section assumes that this gas has been properly connected and the
line checked for leaks.
To initiate a span calibration:
<UP/DOWN >From the MAIN MENU, scroll down to the SPAN
function.
<Enter>
Press <Enter> to activate the SPAN function.
________________________________________________
________________________________________________
The first line allows the user to modify the span target value.
The second line commences the span after the following query:
Teledyne Analytical Instruments
50
: Determines compound names, associated
Logging out prevents unauthorized tampering with
: Removes power to outputs and displays, but
-> Span 80.00 ppb
Span Begin
Series 4060