Alarm Relays; Table 3-3: Analog Range Id Output-Example - Teledyne 4060 Series Instruction, Operating And Maintenance Manual

Trace aromatic hydrocarbon analyzer
Table of Contents

Advertisement

Installation
40
0.4
50
0.5
60
0.6
70
0.7
80
0.8
90
0.9
100
1.0
Table 3-3: Analog Range ID Output—Example
Range
Voltage Signal
Output (VDC)
Range 1
0.20
Range 2
0.50
Range 3
0.80
3.3.2.5 A
LARM
The nine alarm-circuit connector pins connect to the internal alarm
relay contacts. Each set of three pins provides one set of Form C relay
contacts. Each relay has both normally open and normally closed contact
connections. The contact connections are shown in Table 3-4. They are
capable of switching up to 3 amperes at 250 VAC into a resistive load.
The connectors are:
Threshold Alarm 1:
 Can be configured as high (actuates when
concentration is above threshold), or low (actuates
when concentration is below threshold).
 Can be configured as failsafe or non-failsafe
 Can be configured as non-latching
 Can be configured out (defeated).
Teledyne Analytical Instruments
32
10.4
12.0
13.6
15.2
16.8
18.4
20.0
Current Signal
Output (mA DC)
7.2
12
16.8
R
ELAYS
Series 4060
Application
0–100 ppb Benzene
0–1 ppm
Benzene
0–10 ppm Benzene