Installation
40
0.4
50
0.5
60
0.6
70
0.7
80
0.8
90
0.9
100
1.0
Table 3-3: Analog Range ID Output—Example
Range
Voltage Signal
Output (VDC)
Range 1
0.20
Range 2
0.50
Range 3
0.80
3.3.2.5 A
LARM
The nine alarm-circuit connector pins connect to the internal alarm
relay contacts. Each set of three pins provides one set of Form C relay
contacts. Each relay has both normally open and normally closed contact
connections. The contact connections are shown in Table 3-4. They are
capable of switching up to 3 amperes at 250 VAC into a resistive load.
The connectors are:
Threshold Alarm 1:
Can be configured as high (actuates when
concentration is above threshold), or low (actuates
when concentration is below threshold).
Can be configured as failsafe or non-failsafe
Can be configured as non-latching
Can be configured out (defeated).
Teledyne Analytical Instruments
32
10.4
12.0
13.6
15.2
16.8
18.4
20.0
Current Signal
Output (mA DC)
7.2
12
16.8
R
ELAYS
Series 4060
Application
0–100 ppb Benzene
0–1 ppm
Benzene
0–10 ppm Benzene