Sample Cell; Detector Module - Teledyne 6000B Operating Instructions Manual

Photometric analyzer
Table of Contents

Advertisement

Photometric Analyzer

1.6.2 Sample Cell

The sample cell rests in a module placed between the source and detector
module. The module contains the sample cell and optional heater and
thermistor for temperature-controlled sample cells.

1.6.3 Detector Module

The detector module houses the photo detector, chopper assembly, and
the signal processing stages of the electronics circuitry. The synchronized
chopper motor rotates at 1800 rpm. The detector type found in your ana-
lyzer can be identified from the letter in the model number (either B or P).
The filter wheel that carries the optical filters is marked with (M) for
measuring and R for reference filter. If you remove the filter wheel, you must
align a reference mark on the wheel with a reference mark on the shaft. When
the switch activating disc is removed, align with the marks on the switch plate
and motor mount when you put it back.
The phototube detector PC board contains the I to E converter stage,
second amplifier, logarithmic amplifier, inverter, and first stage of integration.
The solid state detector has its I to E converter stage built in on the detector PC
board. A system with a solid state detector has a second converter PC board
containing the second amplifier, logarithmic amplifier, inverter, and first stage
of integration.
The magnetically-activated reed switch is mounted on the motor
mount. Oscilloscope test points are available and are mounted on a bracket
inside the housing for explosion-proof models; test points are available on the
outside in the bottom for general-purpose units. An optional zero and/or span
filter is located in this module also.
Exposed Sample Module
Teledyne Analytical Instruments
Operational Theory 1
Part II: