2 Installation
Primary Input Power:
accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed
connections.
DANGER: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The standby function switches power on or
off to the displays and outputs only.
Figure 2-4: Primary Input Power Connections
Fuse Installation:
type (slow blow) fuses. Fuses are not installed at the factory. Be sure to
install the proper fuse as part of installation (See Fuse Replacement in
chapter 4, maintenance.)
Analog Outputs:
the ANALOG OUTPUTS terminal block. There are two connectors per
output with the polarity noted. See Figure 2-5.
The outputs are:
0–1 V dc % of Range: Voltage rises linearly with increasing sample con-
0–1 V dc Range ID:
2-4: Part I
The power supply in the Model 6000B will
115VAC
The fuse holders accept 5 x 20 mm, 4.0 A, T
There are eight DC output signal connectors on
centration, from 0 V at 0% to 1 V at 100%. (Full
scale = 100% programmed range.)
0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Cal Range.