Trace Oxygen Analyzer
and Interconnection Drawings in the Drawings section of this manual. It
also provides other pertinent information applicable to your instrument.
For special applications the micro-fuel cell may also be of a
different type than the standard A-2C, B-2C or Z-2C unit. If this is the
case, the pertinent cell specifications will be given in the addendum.
3.2.3 Installing the Micro-fuel Cell/Cell Block Orientation
A micro-fuel cell is included as a separate item. It must be installed
prior to instrument use.
Also, once it is expended, or if the instrument has been idle for a
lengthy period, the micro-fuel cell will need to be replaced.
Figure 3-1: Micro-fuel Cell Orientation
During the Installation and/or Replacement of the MFC, Membrane
surface MUST ALWAYS FACE DOWNWARD, and the Contact side of
the Membrane, MUST be placed FIRST into analysis unit.
The reason for proper Installation/Replacement is, if any bubble
that develops as the electrolyte dries out will be directed by gravity away
from the membrane.
To install or replace the micro-fuel cell, follow the procedures in
Chapter 5, Maintenance.