Manual Mode Zeroing - Teledyne 6000B Operating Instructions Manual

Photometric analyzer
Table of Contents

Advertisement

Photometric Analyzer
the lower right corner of the screen as C—Zero, for 3 min, and then does a fine
zero, and displays F—Zero, for 3 min.
Then, and whenever Slope is less than 0.01 for at least 3 min, instead of
Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are
software steps in the zeroing process that the system must complete, AFTER
settling, before it can go back to Analyze. Software zero is indicated by S–
Zero in the lower right corner.
The zeroing process will automatically conclude when the output is within
the acceptable range for a good zero. Then the analyzer automatically returns to
the Analyze mode.
Press Zero to enter the Zero function. The screen that appears allows you
to select between automatic or manual zero calibration. Use the Δ∇ keys to
toggle between AUTO and MAN zero settling. Stop when MANUAL appears,
blinking, on the display.
Press Enter to begin the zero calibration. After a few seconds the first of
three zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predeter-
mined rate.
The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–Zero
and F–Zero, use the Δ Δ Δ Δ Δ ∇ keys to adjust displayed Zero adj: value as close as
possible to zero. Then, press Enter.
S–Zero starts. During S–Zero, the Microcontroller takes control as in Auto
Mode Zeroing, above. It calculates the differences between successive sam-
plings and displays the rate of change as Slope= a value in parts per million per
second (ppm/s).
####.##
4 Left=#.###
3.4.1.2

Manual Mode Zeroing

Select zero
mode:
####.##
Zero adj:2048 C—Zero
####.## % SO2
Slope=#.### S—Zero
Teledyne Analytical Instruments
%
SO2
S—Zero
MANUAL
% SO2
Operation 3
Part I
3-17

Advertisement

Table of Contents
loading

Table of Contents