4-25. Fault Codes

100
Low Air Flow
200
5725 No Error
201
5725 Self-Test ROM Failure
202
5725 Self-Test RAM Failure
203
5725 Self-Test EEPROM Failure
204
5725 Self-Test Data Bus Failure
205
5725 Self-Test CLAMPS Circuit Failure
206
5725 Self-Test HVCLR Circuit Failure
207
5725 Self-Test DAC Failure
208
5725 Self-Test Watchdog Timer Failure
209
5725 Current Heatsink Too Hot
210
Output Tripped To Standby
211
5725 Current Compliance Voltage Too High
212
5725 Current Compliance Voltage Too High
213
5725 +400V Supply Did Not Shut Off
214
5725 -400V Supply Did Not Shut Off
215
5725 Voltage Heatsink Too Hot
216
5725 Voltage Heatsink Too Hot
217
5725 +400V Supply Too Small
218
5725 +400V Supply Too Large
219
5725 -400V Supply Too Large
220
5725 -400V Supply Too Small
221
5725 +400V Supply Current Too High
222
Output Tripped To Standby
223
5725 -400V Supply Current Too High
224
Output Tripped To Standby
225
5725 Fan Not Working
226
5725 CLAMPS Fault
227
Output Tripped To Standby
228
5725 Software TRAP
229
5725 Cable Was Off
100-Level Faults: Air Flow Monitoring Subsystem
200-Level Faults: 5725 Boost System
Maintenance
4

Fault Codes