5700A/5720A Series II Calibrator
Service Manual
3-6.

Calibration Procedure

Before you being this procedure make sure the calibrator is powered on and has
completed the appropriate warm-up period.
After you have finished calibration, but before you save the new constants, the calibrator
presents the new changes as ± ppm, and as a percentage of specifications for each range
and function. A list of changes can be sent to a computer through either the serial or
instrument control (IEEE-488) port, or printed through the serial port. The largest
proposed change will be displayed on the calibrator's front panel.
Follow this procedure to calibrate the main output functions.
1. Press the "Setup Menus" softkey; then press the "Cal" softkey. The calibration menu
2. Press the "Cal" softkey. The display shows the following.
3. To calibrate the main output functions, press one of the softkeys under "Calibration."
3-6
appears as shown below:
PREV
MENU
PREV
MENU
The display shows the following:
PREV
MENU
IF3-1.EPS
IF3-3.EPS