Fluke 5700A Series II Service Manual page 127

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

Output
Frequency
Level
200V
1 kHz
200V
20 kHz
200V
50 kHz
200V
100 kHz
200V(2)
40 Hz
200V
20 Hz
200V
10 Hz
300V(3)
20 kHz
600V(3)
50 kHz
600V(3)
100 kHz
1 kV
1 kHz
1 kV
50 Hz
1 kV
300 Hz
1 kV(3)
20 kHz
1 kV(3)
30 kHz
1 kV(3)
40 Hz
1: This is a test of the bottom of the 20V range .
2: Observe the Calibrator output for 10 minutes and verify that it remains stable within 7.5 ppm.
3: Perform only for units that are used with a 5725A Amplifier.
4: When the TUR (Test Uncertainty Ratio) is less than 4:1, the spec is reduced to give the same
Consumer Risk as a 4:1 TUR as described later in this chapter.
Table 3-23. 5720A AC Voltage Test Record (cont)
Error
Spec
Display
( ppm )
Reading
64
64
101
185
64
110
295
145
378
1375
79
79
79
131
371
84
Determining Test Limits for Other Calibration Intervals
5790A
1-Year
TUR
Spec
Factor
( ppm)
31
2.05
0.936
31
2.05
0.936
69
1.46
0.875
98
1.89
0.923
31
2.05
0.936
68
1.62
0.896
200
1.48
0.878
41
3.54
.993
130
2.91
.977
500
2.21
.945
38
2.08
0.938
38
2.08
0.938
38
2.08
0.938
38
3.45
.991
130
2.85
.975
38
2.21
.945
Calibration and Verification
GB
Test Limit (4)
( ppm)
59
59
88
171
59
99
259
144
369
1299
74
74
74
130
362
79
3
3-57