Fluke 5700A Series II Service Manual page 137

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

Div. Setting
5700A/5720A Series II
10V
100V
1000V
Kelvin-Varley Setting
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
Differential Amplifier
Sensitivity
50 V/division
50 V/division
50 V/division
100 V/division
500 V/division
Table 3-34. Load Regulation Test Record
5700A/5720A Series II
Out/Full Load
Range
11V
10V/199
220V
100V/1999
1100V
1000V/49.99 k
Table 3-35. DC Voltage Linearity Test Record
5700A/5720A Series II
Output
1V
2V
3V
4V
5V
6V
7V
8V
9V
Table 3-36. DC Voltage Output Noise Test
Calibrator (UUT)
2.2V
10V
20V
200V
1000V
Determining Test Limits for Other Calibration Intervals
Change In Null
Measured
Null Detector
Reading
Rms Meter
Reading
Calibration and Verification
Change In
Null Limit
3.1 V
42 V
310 V
Null Detector
Reading limit
2.3 V
2.6 V
2.9 V
3.2 V
3.5 V
3.8 V
4.1 V
4.4 V
4.7 V
Maximum rms Meter
Reading
150 mV
500 mV
500 mV
750 mV
500 mV
3
3-67