Fluke 5700A Series II Service Manual page 31

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

Range
Resolution
99 % Confidence Level
220 mV
10 nV
2.2 V
100 nV
11 V
1 V
22 V
1 V
220 V
10 V
1100 V
100 V
95 % Confidence Level
220 mV
10 nV
2.2 V
100 nV
11 V
1 µV
22 V
1 µV
220 V
10 µV
1100 V
100 µV
Notes:
DC Zeros calibration required every 30 days.
1.
For fields strengths >1 V/m but
5700A Series II DC Voltage Specifications
Absolute Uncertainty
5 °C from calibration temperature
24 Hours
90 Days
6.5 + .75
7 + .75
3.5 + 1.2
6 + 1.2
3.5 + 3
5 + 4
3.5 + 6
5 + 8
5 + 100
6 + 100
7 + 600
8 + 600
5.5 + 0.6
6 + 0.6
3.5 + 1
5 + 1
3 + 3.5
4 + 3.5
3 + 6.5
4 + 6.5
4 + 80
5 + 80
6 + 500
7 + 500
3 V/m, add 0.01 % of range.
Introduction and Specifications
[1]
180 Days
1 Year
24 Hours
(ppm output + V)
8 + .75
9 + .8
2.5 + .5
7 + 1.2
8 + 1.2
2.5 + 1.2
7 + 4
8 + 4
7 + 8
8 + 8
8 + 100
9 + 100
2.5 + 100
10 + 600
11 + 600
3 + 600
7 + 0.6
8 + 0.6
6 + 1
7 + 1
6 + 3.5
7 + 3.5
6 + 6.5
7 + 6.5
7 + 80
8 + 80
8 + 500
9 + 500
2.4 + 500
Electrical Specifications
Relative Uncertainty
1 °C
90 Days
4 + .5
4 + 1.2
1.5 + 3
3.5 + 4
1.5 + 6
3.5 + 8
4 + 100
4.5 + 600
2 + 0.4
3.5 + 0.4
2 + 1
3.5 + 1
1.2 + 3
3 + 3.5
1.2 + 6
3 + 7
2 + 80
3.5 + 80
4 + 500
1
1-15