1-5. Wideband AC Voltage Module (Option 5700A-03)

The Wideband AC Voltage Module (Option 5700A-03) can be installed in both the
5700A and 5720A Series II Calibrators. The module is a high-accuracy, low-noise,
extremely flat ac voltage source for calibrating rf voltmeters, with a frequency range of
10 Hz to 30 MHz. Output is in seven ranges from 300 µV (-57 dBm) to 3.5 V (+24 dBm)
through a Type-N coaxial connector into a 50
or dBm through either the front panel controls or under remote control.
The wideband module also functions with the calibrator's output adjust controls that let
display the error of a wideband meter in either percentage of output or in decibels.
Included with the wideband module is a Type-N output cable and a 50
wideband module is calibrated to the end of its standard-equipment output cable.
Introduction and Specifications

Wideband AC Voltage Module (Option 5700A-03)

load. The output level is selected in volts
terminator. The