1-28. Electrical Specifications

Fluke guarantees performance verification using specifications stated to
99% confidence level.

1-29. DC Voltage Specifications

Range
99 % Confidence Level
220 mV
2.2 V
11 V
22 V
220 V
1100 V
95 % Confidence Level
220 mV
2.2 V
11 V
22 V
220 V
1100 V
Notes:
DC Zeros calibration required every 30 days.
1.
For fields strengths >1 V/m but 3 V/m, add 0.01 % of range.
1-14
5720A Series II DC Voltage Specifications
5 °C from calibration temperature
Resolution
24 Hours
10 nV
5 + 0.5
100 nV
3.5 + 0.8
2.5 + 3
1 V
2.5 + 5
1 V
3.5 + 50
10 V
5 + 500
100 V
10 nV
4 + 0.4
100 nV
3 + 0.7
2 + 2.5
1 V
2 + 4
1 V
3 + 40
10 V
4 + 400
100 V
Note
Absolute Uncertainty
90 Days
180 Days
1 Year
(ppm output + V)
7 + 0.5
8 + 0.5
9 + 0.5
4 + 0.8
4.5 + 0.8
6 + 0.8
3 + 3
3.5 + 3
3 + 5
3.5 + 5
4 + 50
5 + 50
6 + 50
6 + 500
7 + 500
8 + 500
6 + 0.4
6.5 + 0.4
7.5 + 0.4
3.5 + 0.7
4 + 0.7
5 + 0.7
2.5 + 2.5
3 + 2.5
3.5 + 2.5
2.5 + 4
3 + 4
3.5 + 4
3.5 + 40
4 + 40
5 + 40
4.5 + 400
6 + 400
6.5 + 400
Relative Uncertainty
[1]
1 °C
24 Hours
2 + 0.4
2 + 0.8
4 + 3
1 + 3
4 + 5
1 + 5
2 + 50
2.5 + 400
1.6 + 0.4
1.6 + 0.7
0.8 + 2.5
0.8 + 4
1.6 + 40
2 + 400
90 Days
2.5 + 0.4
2.5 + 0.8
1.5 + 3
1.5 + 5
2.5 + 50
3 + 400
2 + 0.4
2 + 0.7
1.2 + 2.5
1.2 + 4
2 + 40
2.4 + 400