Table 3-21. Direct Current Accuracy Test Record (5700A)
Std
Output
DC DMM
Res
Current
Setting
Value
+2A
0.1
(200 mV)
-2A
0.1
(200 mV)
+200 mA
1.0
(200 mV)
-200 mA
1.0
(200 mV)
+20 mA
10
(200 mV)
-20 mA
10
(200 mV)
+2 mA
1 k
(2V)
-2 mA
1 k
(2V)
+200 A
10 k
(2V)
-200
A
10 k
(2V)
Output
DC DMM
Current
Reading
+10A
-10A
+5A
+3A
-3A
Table 3-22. AC Voltage Frequency Accuracy Test Record
Frequency
10 Hz
15 Hz
100 Hz
200 Hz
500 Hz
1 kHz
5 kHz
10 kHz
140 kHz
200 kHz
500 kHz
1 MHz
Characterizing dc DMM @ 200 mV and 2.0V
Column A
Rdg
Std Res
Correction
(ppm)
5725A Amplifier DC Current Test
Certified Shunt
Value
Tolerance
99.99 ms - 100.01 ms
66.673 ms - 66.66 ms
9.999 ms - 10.001 ms
199.98 Hz - 200.02 Hz
499.95 Hz - 500.05 Hz
999.9 Hz - 1000.1 Hz
4999.5 Hz - 5000.5 Hz
9.999 kHz - 10.001 kHz
139.986 kHz - 140.014 kHz
199.98 kHz - 200.02 kHz
499.95 kHz - 500.05 kHz
0.9999 MHz - 1.0001 MHz
Determining Test Limits for Other Calibration Intervals
Column B
UUT Actual
UUT Error
Error (A+B)
Reading
(ppm)
(ppm)
Cal Actual
Current
Calibration and Verification
Test Limit,
90-day
(ppm)
135
135
73
73
55
55
55
55
100
100
Test Limits, 90-
day
( ppm)
388 ppm
388 ppm
436 ppm
500 ppm
500 ppm
Actual
3
3-55