Resistance Sourcing

Range
11.000
110.00
1.1000 k
11.000 k
Temperature Coefficient: (0.01% of f.s.)/ C in the ranges -10 to 18 C and 28 to 50 C
Maximum and Minimum Current through Source Resistance:
11
Range: 3 mA dc max, 0.1 mA dc min
110
Range: 3 mA dc max, 0.1 mA dc min
1.1 k
Range: 3 mA dc max, 0.01 mA dc min
11 k
Range: 1 mA dc max, 0.01 mA dc min
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc
Resolution
1 m
10 m
100 m
1
Documenting Process Calibrator
Models 741B and 743B
% of Output + ohms
1 Year
0.01% + 0.02
0.01% + 0.04
0.02% + 0. 5
0.03% + 5
Specifications
2 Year
0.02% + 0.02
0.02% + 0.04
0.03% + 0.5
0.04% + 5