Fluke 741B User Manual page 88

Documenting process calibrator
Hide thumbs Also See for 741B:
Table of Contents

Advertisement

741B/743B
Users Manual
7. Press
for SOURCE mode.
8. Press the appropriate function key for the control
output (e.g.,
or
connected to a current loop that has a power
supply, select
Simulate Transmitter
current output choice.
9. Select a source value, e.g., 4 mA.
10. Press
for MEASURE/SOURCE mode.
11. Press
More Choices
softkey appears.
12. Press the
Transmitter Mode
13. Set the 0% and 100% values for MEASURE and
SOURCE on the display. You can select
or
for the transfer function.
14. Press
.
Done
78
). If the transmitter is
for the
until the
Transmitter Mode
softkey.
Linear
15. The calibrator is now in Transmitter mode. It is
measuring the process input and sourcing the
control signal output proportional to the input.
16. To change any of the Transmitter mode
parameters, press
Change Setup
process in step 13.
17. To exit Transmitter mode, press the
softkey.
, and repeat the