As Left" Test Run - Fluke 741B User Manual

Documenting process calibrator
Hide thumbs Also See for 741B:
Table of Contents

Advertisement

741B/743B
Users Manual
4. Adjust the transmitter output for 20 mA.
5. If the span was adjusted in step 4, you must go
back and repeat steps 3 and 4 until no more
adjustment is required.
6. Now check the transmitter at 50%. If it is within
specification, your adjustment is complete. If not,
adjust the linearity and begin this procedure
again at step 3.

"As Left" Test Run

Proceed as follows to generate and record as left
data for the thermocouple temperature transmitter
you have just adjusted.
1. Press the
softkey to record as left data.
As Left
2. Press the
Auto Test
automatic sequence through all the test points,
or you can step through the tests manually.
3. When the tests are complete, observe the error
summary table, such as the following.
72
softkey to begin an
An asterisk (
) next to a measure or source value
*
indicates an unsettled value (
when the measurement was taken.
4. If all the results are within specification, as they
are this time, press the
memory is made for as left data.
gj48s.eps
g
annunciator)
softkey. An entry in
Done