741B/743B
Users Manual
80T-150U Temperature Probe
80PK series thermocouples
80i-410 Clamp-on DC/AC Current Probe
80i-1010 Clamp-on DC/AC Current Probe
80i-500s Clamp-on AC Current Probe (requires
the Y9108 adapter)
80i-1000s Clamp-on AC Current Probe (requires
the Y9108 adapter)
80i-kW Current and Power Probe

Specifications

All specifications apply from +18 C to +28 C unless
stated otherwise.
All specifications assume a 5 minute warmup period.
102
Measurement specifications are valid only when
Damping is turned on. When damping is turned off,
g
or when the
annunciator is displayed, floor
specifications are multiplied by 3. Floor specifications
are the second part of the specifications, usually
expressed as "% of full scale." The measure
pressure, temperature, and frequency functions are
specified only with damping on.
The standard specification intervals for the 741B and
743B are 1 and 2 years. Typical 90-day source and
measurement accuracy can be estimated by dividing
the 1 year "% of Reading" or "% of Output"
specifications by 2. Floor specifications, expressed
as "% of f.s.", remain constant.
To achieve the best noise rejection, use battery
power and tie all three common jacks together.