Dc Current Source; Dc Current Simulate (External Loop Power); Resistance Sourcing - Fluke 753 User Manual

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

753/754
Users Manual

+DC Current Source

Range/Mode
0.100 to 22.000 mA
Temperature Coefficient: 3 % of specified accuracy / C (<18 C or >28 C)
Source mA Compliance Voltage: 18 V maximum
Source mA Open Circuit Voltage: 30 V maximum

+DC Current Simulate (External Loop Power)

Range/Mode
0.100 to 22.000 mA (Current Sink)
Simulate mA Input Voltage: 15 to 50 V dc, add 300 A to floor when >25 V is present on the loop
Temperature Coefficient: 3 % of specified accuracy / C (<18 C or >28 C)

Resistance Sourcing

Range
10.000
0.001
100.00
0.01
1.0000 k
0.1
10.000 k
1
Temperature Coefficient: (0.01 % of output +0.02 % of range / C (<18 C or >28 C)
110
Resolution
1 A
Resolution
1 A
Resolution
0.01 % + 0.010
0.01 % + 0.02
0.02 % + 0.0002 k
0.02 % + 0.003 k
% of Output + Floor
1 Year
0.01 % + 3 A
% of Output + Floor
1 Year
0.02 % + 7 A
% of Output + Floor
1 Year
2 Year
0.015 % + 0.015
0.015 % + 0.03
0.03 % + 0.0003 k
0.03 % + 0.005 k
2 Year
0.02 % + 3 A
2 Year
0.04 % + 7 A
Allowable Excitation
Current
0.1 mA to 10 mA
0.1 mA to 10 mA
0.01 mA to 1.0 mA
0.01 mA to 1.0 mA

Advertisement

Table of Contents
loading

This manual is also suitable for:

754

Table of Contents