Using Auto Step - Fluke 741B User Manual

Documenting process calibrator
Hide thumbs Also See for 741B:
Table of Contents

Advertisement

Using Auto Step

To have the calibrator make a sequence of steps
automatically, either once through the sequence or
repetitively, proceed as follows:
1. Refer to the appropriate "Using Source Mode"
subheading earlier in this manual (e.g.,
"Sourcing Electrical Parameters") and connect
the calibrator to the circuit to be tested.
2. If necessary, press
3. Set the calibrator for the desired source value.
4. If you want to step the source value in % of
scale, set the % of scale value as described
previously under "Sourcing in Percent of Scale."
5. Press the
softkey.
Step
6. Press the
Auto Step
for SOURCE mode.
softkey.
Documenting Process Calibrator
7. Select values for the following parameters as
you are prompted by the display:
Starting point (in units or % of scale)
Ending point
Number of steps
Time per step
Single shot or continuous repetition
Ramp pattern if continuous
Start delay
8. To start automatic stepping, press the
softkey. The softkey label changes to
9. To stop automatic stepping, press the
softkey.
10. Press the
softkey to resume normal
Done
operation.
Using Source Mode
Start Step
.
Stop Step
Stop Step
59