741B/743B
Users Manual
To measure temperature using an RTD input,
proceed as follows:
1. If necessary, press
2. Press
twice so that
3. Press
or
to select the desired RTD type.
4. Press
.
5. Press
or
to select a 2-, 3-, or 4- wire
connection.
6. Attach the RTD to input jacks as the display or
Figure 15 shows. Use the supplied jumper
between the mA
RTD MEAS low jack and the
V MEAS low jack as shown if you are using a 3-
wire connection.
7. Press
.
38
for MEASURE mode.
shows.
Select RTD Type
Caution
Do not force a dual banana plug between
any two jacks in the horizontal
orientation. Doing so will damage the
jacks. Use the supplied jumper wire when
needed for RTD measurements. You can
use a dual banana plug in the vertical
orientation. See Figure 14.