Using Resistance-Temperature Detectors (Rtds); Rtd Types Accepted - Fluke 741B User Manual

Documenting process calibrator
Hide thumbs Also See for 741B:
Table of Contents

Advertisement

Using Resistance-Temperature Detectors
(RTDs)
The calibrator accepts RTD types shown in Table 6.
RTDs are characterized by their resistance at 0°C
(32°F), which is called the "ice point" or R
common R
is 100 . Most RTDs come in a three-
0
RTD Type
Ice Point (R
Pt100 (3926)
*Pt100 (385)
Ni120 (672)
Pt200 (385)
Pt500 (385)
Pt1000 (385)
Cu10 (427)
Pt100 (3916)
*Per IEC 751-Standard
. The most
0
Table 6. RTD Types Accepted
)
Material
0
100
Platinum
100
Platinum
120
Nickel
200
Platinum
500
Platinum
1000
Platinum
9.035
**
Copper
100
Platinum
**10
@ 25 C
Documenting Process Calibrator
terminal configuration. The calibrator accepts RTD
measurement inputs in two-, three-, or four-wire
connections as shown in Figure 15. A four-wire
configuration provides the highest measurement
precision, and two-wire provides the lowest
measurement precision.
0.003926 / C
0.00385 / C
0.00672 / C
0.00385 / C
0.00385 / C
0.00385 / C
0.00427 / C
0.003916 / C
Using Measure Mode
Range ( C)
-200 to 630
-200 to 800
-80 to 260
-200 to 630
-200 to 630
-200 to 630
-100 to 260
-200 to 630
37