Calibrate:state; Calibrate:voltage; Calibrate:voltage:protection - Agilent Technologies 66312A Programming Manual

Dynamic measurement dc source; system dc power supply
Hide thumbs Also See for 66312A:
Table of Contents

Advertisement

4 - Language Dictionary

CALibrate:STATe

This command enables and disables calibration mode. The calibration mode must be enabled before the
will accept any other calibration commands.
The first parameter specifies the enabled or disabled state. The second parameter is the password. It is
required if the calibration mode is being enabled and the existing password is not 0. If the password is not
entered or is incorrect, an error is generated and the calibration mode remains disabled. The query
statement returns only the state, not the password.
NOTE:
Whenever the calibration state is changed from enabled to disabled, any new calibration
constants are lost unless they have been stored with CALibrate:SAVE.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands

CALibrate:VOLTage

This command initiates the calibration of the output voltage and the voltage readback circuit.
Command Syntax
Parameters
Examples

CALibrate:VOLTage:PROTection

This command can calibrates the overvoltage protection (OV) circuit. The dc source automatically
performs the calibration. CALibrate:VOLTage:PROTection is a sequential command that takes several
seconds to complete.
Command Syntax
Parameters
Examples
46
CALibrate:STATe<bool>[,<NRf>]
0 | 1 | OFF | ON [,<password>]
OFF
CAL:STAT 1,6812
CAL:STAT OFF
CALibrate:STATe?
<NR1>
CAL:PASS
CAL:SAVE
CALibrate:VOLTage[:DC]
None
CAL:VOLT
CAL:VOLT:DC
CALibrate:VOLTage:PROTection
None
CAL:VOLT:PROT
*RST

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6631b6632b66332a6633b6634b6611c ... Show all

Table of Contents