Measuring Output Pulses (Agilent 66312A, 66332A Only); Current Detector; Pulse Measurement Queries - Agilent Technologies 66312A Programming Manual

Dynamic measurement dc source; system dc power supply
Hide thumbs Also See for 66312A:
Table of Contents

Advertisement

3 - Programming the DC Source

Measuring Output Pulses (Agilent 66312A, 66332A Only)

Current Detector

Check that the current detector is set to ACDC when measuring current pulses or other waveforms with a
frequency content greater than a few kilohertz.
SENSe:CURRent:DETect ACDC
Only select DC as the measurement detector if you are making only DC current measurements and you
require a measurement offset better than 2mA on the High current measurement range. Note that this
selection gives inaccurate results on current waveforms that have ac content.
SENSe:CURRent:DETect DC

Pulse Measurement Queries

The dc source has several measurement queries that return key parameters of pulsewaveforms as shown
in Figure 3-4.
FETC:CURR:MAX?
FETC:VOLT:MAX?
FETC:CURR:HIGH?
FETC:VOLT:HIGH?
DATA POINTS
Figure 3-4. Measurement Commands Used to Return Pulse Data
To return the maximum or minimum value of a pulse waveform use the following commands. Note that
the data points of the measurement sample may not coincide with the actual maximum or minimum point
on the waveform.
FETCh:VOLTage:MAXimum? or
FETCh:VOLTage:MINimum?
FETCh:CURRent:MAXimum? or
FETCh:CURRent:MINimum?
The average value of the high level or low level of a pulse can also be measured. To return the average
value of the high level, use:
FETCh:CURRent:HIGH? or
FETCh:VOLTage:HIGH?
To return the average value of the low level, use:
FETCh:CURRent:LOW?
FETCh:VOLTage:LOW?
28
FETC:CURR:LOW?
FETC:VOLT:LOW?
FETC:CURR:MIN?
FETC:VOLT:MIN?
or

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6631b6632b66332a6633b6634b6611c ... Show all

Table of Contents