4 - Language Dictionary
* OPC does not prevent processing of subsequent commands, but bit 0 will not be set until all pending
operations are completed.
*OPC? causes the instrument to place an ASCII "1" in the Output Queue when all pending operations are
completed. Unlike *OPC, *OPC? prevents processing of all subsequent commands. It is intended to be
used at the end of a command line so that the application program can then monitor the bus for data until
it receives the "1" from the dc source Output Queue.
Command Syntax
Parameters
Query Syntax
Returned Parameters
Related Commands

*PSC

This command controls the automatic clearing at power-on of the Service Request Enable and the
Standard Event Status Enable registers
causes these registers to be cleared at power-on. This prevents a PON event from
*PSC ON | 1
generating SRQ at power-on.
causes the contents of the Standard Event Enable and Service Request Enable registers
*PSC OFF | 0
to be saved in nonvolatile memory and recalled at power-on. This allows a PON event to
generate SRQ at power-on.
The query returns the current state of *PSC.
Command Syntax
Parameters
Example
Query Syntax
Returned Parameters
Related Commands
CAUTION:
*PSC causes a write cycle to nonvolatile memory. Nonvolatile memory has a finite
maximum number of write cycles. Programs that repeatedly cause write cycles to
nonvolatile memory can eventually exceed the maximum number of write cycles and
cause the memory to fail.

*SRE

This command sets the condition of the Service Request Enable Register. This register determines which
bits from the Status Byte Register (see *STB for its bit configuration) are allowed to set the Master Status
Summary (MSS) bit and the Request for Service (RQS) summary bit. A 1 in any Service Request Enable
Register bit position enables the corresponding Status Byte Register bit and all such enabled bits then are
logically ORed to cause Bit 6 of the Status Byte Register to be set.
When the controller conducts a serial poll in response to SRQ, the RQS bit is cleared, but the MSS bit is
not. When *SRE is cleared (by programming it with 0), the dc source cannot generate an SRQ to the
controller.
The query returns the current state of *SRE.
66
*OPC
None
*OPC?
<NR1> 1
*OPC
*TRIG
*WAI
*PSC <Bool>
0 | 1 | OFF | ON
*PSC 0
*PSC 1
*PSC?
<NR1>0|1
*ESE
*SRE