Setting Thresholds
In this mode, set the measured values for a PASS (OK) judgment. Both HIGH and LOW
threshold values are set. Three judgment results are output; "HIGH", "PASS" and "LOW".
HIGH threshold
Measured value
LOW threshold
ON
HIGH
OFF
Output
ON
PASS
OFF
ON
LOW
OFF
Note
In the special mode (IC lead pitch or IC lead width judgment mode), the following values are
output:
HIGH:Standard value setting
LOW: Tolerance setting
Measurement Cycle p.61
The following describes, as an example, the operation procedure for setting a HIGH threshold.
1
Set the value.
Move from one digit to another.
Change the current value.
Confirm the selection.
Note
Hysteresis can also be set for threshold values. Set hysteresis when judgments are unstable.
Hysteresis p.67
ZX-GT User's Manual
- T Mode
(ON when measured value > HIGH threshold)
(ON when LOW threshold ≤ measured value
≤ HIGH threshold)
(ON when measured value < LOW threshold)
SUB

Setting Thresholds - T Mode

Measured
value
Threshold
45