Connecting The P-200 Printer - Nellcor N-200 Operator's Manual

Hide thumbs Also See for N-200:
Table of Contents

Advertisement

CONNECTING THE P-200 PRINTER

The P-200 printer may be used to print the contents of the
trend and event memories, as well as to record ongoing
saturation, pulse rate, pulse amplitude, and monitor status.
To use the P-200 with the N-200:
1.
Set the N-200 oximeter DIP switches as follows:
Switches 2, 4, 5, and 8 up
Switches 3 and 7 down
Switches 1 and 6 either up or down
2.
On the bottom of the P-200, set the DIP switches as
follows:
On the row of eight switches, set 1,2, and 6 down; all others
are up
On the row of six switches, set 4, 5, and 6 down; all others
are up
3.
Connect the printer to the N-200 using the interface cable
provided with the P-200 printer.
4.
Turn on the printer and verify proper operation (see the
P-200 operator's manual).
Connecting to Other Instruments
69