Nellcor N-200 Operator's Manual page 92

Hide thumbs Also See for N-200:
Table of Contents

Advertisement

Specifications
Accuracy and Motion Tolerance
Saturation
Without Motion—Adults:
Without Motion—Neonate
With Motion—Adults and Neonates
4,5
Low Perfusion
:
Pulse Rate
5
Without Motion
With Motion
4,5
Low Perfusion
1
Applicability: D-25, D-25L, and N-25 sensors.
2
Definition of Nellcor Standard Motion:
Tapping:
1-2 cm amplitude at an aperiodic* frequency between 1-4 Hz
Rubbing: 1-2 cm amplitude at an aperiodic* frequency between 1-4 Hz
(*Note: "aperiodic" is a randomly changing frequency)
3
Adult specification is shown for D-25 and N-25 sensors with the N-200. Neonate specification
shown for N-25 sensors with the N-200. Validation testing
conducted on adult volunteers. Saturation accuracy will vary by sensor type. Refer to the Sensor
Accuracy Grid.
4
Definition of Low Perfusion:
Low perfusion in the context of oximetry is generally used to mean that the detected signal
modulation (pulsatility) is low at the monitoring site. Low percent modulation is defined as a value
of 1.5% modulation or less for the infrared (IR) signal.
5
1
Specification applies to monitor performance and was validated with Biotek and Nellcor
simulators.
86
1, 2
70 to 100% ±2 digits
3
:
70 to 100% ±3 digits
3
: 70 to 100% ±3 digits
70 to 100% ±2 digits
20 to 250 bpm ±3 digits
normal physiologic range
(55 – 125 bpm) ±5 digits
20 to 250 bpm ±3 digits
of D-25 and N-25 sensors was