Nellcor N-200 Operator's Manual page 63

Hide thumbs Also See for N-200:
Table of Contents

Advertisement

Xn . . . n
Snnn
Rnnn
Pnnn
Y Packet
This is the time marker description indicating the type of
marker and the time it occurred.
Yn
0
1
2
3
4
5
6
7
Dnnnnnnnn = Time/Date (format mmhhDDMM)
Z Packet
This defines the number of data packets, and the maximum
and minimum values in the following trend or event
information. This can be used for scaling purposes.
Zn . . . n
Snnn
snnn
Rnnn
rnnn
Event data point (n . . . n = current index up to
four digits)
Saturation
Pulse rate
Pulse amplitude
< STX > YnDnnnnnnnn < CRLF > < ETX >
Event Type
User-defined event
Alarm limit exceeded
Patient signal lost
Unused
Time mark (on hour for trend, on min for event)
Power turned off
Patient signal acquired
Clock reset
< STX > Zn . . . nSnnnsnnnRnnnrnnn
< CRLF > < ETX >
Starting scale—n . . . n = (number of data
points, or number of data packets minus 1—
the last W or X packet is a throwaway)
MAX saturation
MIN saturation
MAX pulse rate
MIN pulse rate
Guide to Operation
57

Advertisement

Table of Contents
loading

Table of Contents