General System Problems - Nellcor N-200 Operator's Manual

Hide thumbs Also See for N-200:
Table of Contents

Advertisement

Troubleshooting

General System Problems

1. N-200 does not turn on.
2. N-200 operates on AC power but not on battery.
3. Interface/powerbase is connected to AC power but
the N-200Õ s BATT IN USE indicator stays on.
78
Check the AC and N-200 interface/powerbase
connections for proper AC power supply.
Check battery operation. If the battery is discharged,
connect the N-200 to the interface/powerbase. Plug the
power cord into an outlet supplying the appropriate
mains voltage. The interface/powerbase requires a
minimum of 14hours to recharge the N-200 battery
completely.
Check AC fuse.
The battery may be discharged. To recharge the
battery, connect the N-200 to the interface/powerbase.
Plug the power cord into an appropriate AC power
outlet and confirm that the BATT IN USE indicator is
off. A minimum of 14 hours is required to recharge the
battery completely.
The battery pack or battery charger circuit may be
defective, or the battery fuse may be open.
Check the N-200/interface/powerbase connection. The
BATT IN USE indicator turns off to confirm correct
connection.
Replace the power cord, check the connections, or try
another AC outlet.
Check fuse.

Advertisement

Table of Contents
loading