Performing a UDDS
Calibration
Performing a Touchscreen
Calibration
Alterations
40
UDDS calibration is used to calibrate the ultrasonic document detection
system for multifeeds. UDDS calibration is rarely needed. This calibration
should only be performed if there is a feeding or multifeed detection issue.
1. Select Calibrate. The Calibration screen will be displayed.
2. Touch UDDS Calibration to display the UDDS Calibration screen.
3. You will be prompted to place a 20 lb. or 80 g/m
the input elevator in landscape orientation.
4. Press the Start/Resume button. The calibration results will be displayed.
5. When finished, touch Cancel to return to the Diagnostics screen.
6. Touch Close to return to the Ready screen.
Touchscreen calibration is used to verify that the touch points on the
touchscreen are working correctly. The touchscreen is factory-calibrated. Do
not calibrate the touchscreen unless directed to do so by Technical Support.
Only use this option when directed to do so by Kodak Alaris Technical Support.
2
bond sheet of paper in