TRIGger:SLOPe
Specifies whether a trigger occurs when the selected trigger source signal is
positive-with-respect-to or negative-with-respect-to the specified trigger level.
Command Syntax:
TRIGger:SLOPe POSitive|NEGative
Example Statements:
OUTPUT 719;"Trigger:Slop POSITIVE"
OUTPUT 719;"TRIGGER:SLOP POSITIVE"
Query Syntax:
TRIGger:SLOPe?
Return Format:
CHAR
Attribute Summary:
Synchronization Required:
Preset State:
SCPI Compliance:
Description:
Slope does not apply to free run, source, or GPIB trigger types. These trigger types operate
independently of the slope setting. Trigger delay may also be specified with the SWE:TIM:DEL
command.
Trigger type is specified with the TRIG:SOUR command.
Trigger level is specified with the TRIG:LEV command.
The analyzer must be armed before triggering can start a measurement. See ARM:SOUR for more
information.
no
POS
confirmed
Command Reference
command/query
455