Page 4
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
Page 5
Safety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document. Warning denotes a hazard. It calls attention to a procedure WARNING which, if not correctly performed or adhered to, could result in injury or loss of life.
If there are any signs of damage that may have occurred during shipment, or any accessories appear to be damaged or missing, call your nearest Agilent Technologies sales or service office. Refer to “Contacting Agilent”...
Chapter 1 Electrostatic Discharge Protection Electrostatic Discharge Protection Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the network analyzer. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage.
Chapter 1 Installation Use ESD precautions when performing these installation procedures. CAUTION If the instrument contains Options H70, H7L, or HNQ, a jumper is added to NOTE the rear Plate for connecting the 321.4 MHz IF OUT to the desired input (for the above mentioned options).
Page 13
Chapter 1 Installation 6. Add the washer (2190-0102) and nut (2950-0035) to the BNC connector. Torque the BNC connector to 21 in-lb. See left picture in Figure 7. The SMA connector includes the nut and washer. Torque the SMA connector to 10 in-lb. See right picture in Figure Figure 3...
Page 14
Chapter 1 Installation 10. Insert the PC board into an available option slot in the PSA. Refer to Figure Figure 5 PC Board in Option Slot 11. Install the jumper (8120-5066) to the rear plate from the 321.4 MHz IF OUT connector to the H7L IN connector.
(grounded) side of the ac power supply. Shipping Instructions You must always call the Agilent Technologies Instrument Support Center to initiate service before retuning your instrument to a service office. See “Contacting Agilent” on page 11.
Chapter 1 Safety and Regulatory Information Warnings WARNING The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, which if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Chapter 1 Safety and Regulatory Information WARNING If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only. Cautions The CAUTION notice denotes a hazard.
Chapter 1 Safety and Regulatory Information Instrument Markings When you see this symbol on your instrument, you should refer to the instrument’s instruction manual for important information. This symbol indicates hazardous voltages. The laser radiation symbol is marked on products that have a laser output.