Screen:contents - Agilent Technologies 89400 Series Command Reference Manual

Table of Contents

Advertisement

SCReen:CONTents

Specifies what is displayed on the analyzer screen.
Command Syntax:
SCReen:CONTents TRACe|MSTate|ISTate|MMEMory|MEMory|OPTions|TCAPture
Example Statements:
OUTPUT 719;"SCREEN:CONT TCAPTURE"
OUTPUT 719;"screen:cont TRACE"
Query Syntax:
SCReen:CONTents?
Return Format:
CHAR
Attribute Summary:
Synchronization Required:
Preset State:
SCPI Compliance:
Description:
TRACe displays the standard measurement trace/grid.
MSTate displays the measurement state display. This shows the instrument mode, frequency
parameters, RBW/window settings, etc.
ISTate displays the instrument state display. This shows the range/input, trigger, and source
settings.
MMEMory displays the current disk (mass memory) catalog.
MEMory displays the active memory configuration.
OPTions displays the options configuration.
TCAPture displays the time capture statistics.