[SENSe:]CORRection[1|2]:EXTernal[:STATe]
Enables or disables the use of user-defined calibration coefficients.
Command Syntax:
[SENSe:]CORRection[1|2]:EXTernal[:STATe] OFF|0|ON|1
Example Statements:
OUTPUT 719;":CORR:EXT ON"
OUTPUT 719;"sense:corr:external:stat OFF"
Query Syntax:
[SENSe:]CORRection[1|2]:EXTernal[:STATe]?
Return Format:
Integer
Attribute Summary:
Synchronization Required:
Preset State:
SCPI Compliance:
Description:
You can adjust the gain and delay calibration coefficients for each input channel. By default, the
gain of an input channel is one, the delay is zero seconds. You can adjust these values using the
sense:correction:loss:magnitude and sense:correction:edelay commands.
The values that you specify are not used until you send sense:correction:external ON. To return
to the default gain and delay values, send sense:correction:external OFF.
For additional details, see online help for the [System Utility] [more cal setup] [external
calibration] softkey.
274
no
0
instrument-specific
command/query