Replacing boards and components

3D reconstruction PC

Fig. 38:
NOTE
Disconnect the system from the power supply by pulling the line power plug plug out of
the socket.
The syngo software and the backup/restore disk are kept in the monitor trolley logbook.
Unplug all connectors on the PC.
Unscrew the dongle from the printer port of the PC and reinstall it on the new PC (see
(Fig. 38 / p.
Replace the PC.
Reconnect all connectors.
Perform the startup of the 3D reconstruction PC as described in the Startup Instruc-
tions, Chapter 4.
Then insert the backup / restore disk (in the logbook compartment of the SIREMOBIL
Iso-C) into the disk drive of the PC.
Start the syngo service software.
Select the Backup / Restore menu and restore the data.
Perform a function check of the PC.
Perform the IQ quick test.
Check the DICOM functionality of the PC, if applicable.