Programming - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

50
tube current displayed at 110 kV in mA * 45 mGy/min
measured dose rate value in mGy/min
Multiply the tube current to be programmed by 10, because the entry on the service PC
must be made in 1/10mA. Round off the value to the nearest whole number
Example:
5 (mA) * 45 (mGy/min)
50.98 (mGy/min)

Programming

Connect the service PC to the serial interface of board D1.
Deactivate the write protection of the "Parameters" disk.
Insert the "Parameters" disk of the SIREMOBIL system software in drive A of the ser-
vice PC.
Start the system service software on the service PC.
Enter your name and the group in the corresponding text fields.
Log in by selecting the "System" - "Connect" menus and pressing the "Logon" button.
Enter the password and press the "OK" button.
Answer the prompt "Password has been changed. Put new password to unit?" by click-
ing the "Yes" button.
Close the information window "Password expires on 00:00:00" by clicking the "OK" but-
ton.
The program window "Service PC - SIREMOBIL ..." is displayed.
Select the "Adjustment, Parameters" menu.
In the Combo box "Parameter Groups" select "Limitation of skin dose rate (FL)" .
Click "Get from unit".
Enter the value to be programmed in "Value".
Click "Set Value".
Select "Put to unit". Follow the menu guidance on the service PC.
SIREMOBIL Iso-C
Tube current displayed at 110 kV:
Maximum air kerma rate:
Measured dose rate value:
= 4.41 (mA)
SPR2-230.840.01.01.02
08.08
Adjustments / Programs
= tube current to be programmed in mA
5 mA
45 mGy/min
50.98 mGy/min.
Rounded to the nearest whole number: 44
The value to be programmed is 44
Page 50 of 148
CS PS SP
0
© Siemens, 2008
For internal use only