Settings - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

102

Settings

Resolution
The I.I. should remain in the service position.
WARNING
Attach the resolution test to the I.I. input.
Insert 2.1 mm Cu + 60 mm Al at the radiation exit port of the SIREPHOS.
Use the I.I. Quick test, chapter: Resolution for evaluating optical sharpness.
Check the resolution and, if necessary, adjust the wide ring on the optics.
Perform the additional chapters of the I.Q. Quick test after all the work is complete.
Dose rate
Check the dose rate and, if necessary, adjust it.
Checking the camera optics centering at the I.I. output
Fig. 34:
WARNING
SIREMOBIL Iso-C
Replacing boards and components
X-ray radiation!
See
(Safety information / p.
Use radiation protection! Wear a lead apron.
X-ray radiation!
See
(Safety information / p.
Use radiation protection! Wear a lead apron.
SPR2-230.840.01.01.02
08.08
11).
11).
Page 102 of 148
CS PS SP
0
© Siemens, 2008
For internal use only