Siemens SIREMOBIL Iso-C Troubleshooting Manual page 109

Table of Contents

Advertisement

Replacing boards and components
Language
Display of realtime clock
Display of rotated images
Monitor zoom
Monitor split
Monitor text
Maximum image number
Disable dose rate selection
in user setup
Unit serial number
Product name
Default program change-
able
Create external disk
Default LUT
Default Edge
Software backward com-
patible
Configure the following parameters in User Setup for MEMOSKOP:
Operating mode parame-
ters
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date / Time
Hospital Name
© Siemens, 2008
For internal use only
Set for language of the country.
Configure Yes. Select No only upon customer request.
Configure Yes.
Configure No.
Configure Yes.
Configure Yes.
Configure 100.
Set to "No" if equipped with an dose-area product measure-
ment device. The menu "Storage of Dose on disk" and the
menu "Dose on Hardcopy output" are visible in User Setup.
Otherwise set to "Yes".
Program the serial number of the SIREMOBIL. Do not
enter the leading zero.
Program "SIREMOBIL Iso-C".
Default: Yes. Depending on country-specific regulations
configure Yes or No.
Configure No.
Configure 2. Program a different LUT curve only upon cus-
tomer request.
Configure 1. Program a different Edge enhancement value
only upon customer request.
Configure No.
Change the programmed parameters only upon customer
request.
Configure to monitor A.
For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to monitor B.
For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: configure to both monitors.
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Enter the hospital name.
SPR2-230.840.01.01.02
08.08
Page 109 of 148