Page 2
TI 236 / ARTD-002.731.17... The protective conductor resistance must not exceed 0.2 Ohm. Additional note These instructions apply only to connecting a laser camera to a SIREMOBIL Compact. Parts required • Laser camera connection kit...
Page 3
Video socket Threaded bolt Fig. 1 Fig. 2 Fig. 3 Disconnect the SIREMOBIL Compact from power CAUTION prior to beginning the work. Mechanical installation of the video separator device • Install the video separator device (OTV) as indicated in Fig. 1.
Page 4
Route the ends of the wires through the hole drilled for the remote release socket. • Solder the wires to the remote release socket (refer to 1/Fig.3 and 2/Fig.3). • Install the remote release socket. SIREMOBIL Compact Register 3 RXR2-130.033.02 Page 4 of 8 Siemens AG Rev.
Route the second video cable from the output of the optical video separator device to the insulated, built-in video socket and connect it. Siemens AG Register 3 RXR2-130.033.02 Page 5 of 8 SIREMOBIL Compact Medical Engineering Rev. 02 03.97 TD SP 1...
Page 6
This calls up the User Setup menu for the Memoskop. • Select the "C Documentation" sub-menu. • Configure monitor A for a SIREMOBIL Compact with one monitor. • Configure monitor B for a SIREMOBIL Compact with two monitors. • If the system is equipped with an area dose measurement device (Diamentor) , call up the "K Documentation of Dose"...
This opens the User Setup menu for the Memoskop. • Select the "C Documentation" sub-menu. • Configure monitor A for a SIREMOBIL Compact with one monitor. • Configure monitor B for a SIREMOBIL Compact with two monitors. • If the system is equipped with an area dose measurement device (Diamentor), open the "K Documentation of Dose"...
Page 8
Connecting laser camera This page intentionally left blank. SIREMOBIL Compact Register 3 RXR2-130.033.02 Page 8 of 8 Siemens AG Rev. 02 03.97 TD SP 1 Medical Engineering...