Fluke Datapaq Food Tracker User Manual page 59

Table of Contents

Advertisement

The display panel shows the heater's current temperature, factory-set to °C
or °F. Press and hold the  button to toggle the display between the
temperature units and the heater's target temperature.
To heat the humidity sensor:
1. Switch on the conditioner. It will perform a self-test, then display the current
heater temperature.
2. To set the heater's target temperature, hold down the  button while
pressing
or
temperature about 20°C/36°F above the oven temperature; the exact
amount of the excess will depend on the temperatures involved and the
delay before the system can be placed in the oven.
3. If the temperature set is higher than the current temperature, the status
indicator will flash, indicating that the conditioner is heating.
4. Insert the humidity sensor into the heating chamber in the conditioner and
wait for the display panel to show the target temperature; the status
indicator then stops flashing. The conditioner takes about 30 minutes to
reach the maximum temperature of 150°C/300°F.
5. When the conditioner reaches target temperature, leave the sensor in the
heating chamber for a further 5 minutes. The sensor is now ready for use,
and the run should be started as soon as possible.
FOOD TR ACK ER
Humidity sensor conditioner, CS3050.
. To allow for cooling prior to the run, set the target
Humidity Measurements
59