Fluke Datapaq Food Tracker User Manual page 18

Table of Contents

Advertisement

8-channel MultiPaq21 logger, shown from above and below.
1: Stop button, RED – p. 19
2: Battery status LEDs – p. 20
3: Start button, GREEN – p. 19
4: Logger status LEDs – p. 20
The logger is not designed for use without an external barrier in hot-water
applications or in an environment of rapidly changing temperature; the accuracy of
data cannot then be guaranteed. Although able to record at
recommended that, for freezing applications, even above this temperature, a
thermal barrier (TB5011) be used to overcome thermal shock errors.
MultiPaq21 Logger Specifications
Logger Types
DP2162A
6 channels, type T thermocouples.
DP2166A
6 channels, type K thermocouples.
DP2182A
8 channels, type T thermocouples.
Humidity option on one channel, and RF telemetry option available.
DP2186A
8 channels, type K thermocouples.
Humidity option on one channel, and RF telemetry option available.
Accuracy
Resolution
Measurement range
Operating temperature
(without thermal barrier)
Humidity option
18
Hardware
5: Transmitter aerial socket
6: Thermocouple sockets – p. 43
7: Humidity sensor socket – p. 55
8: Communications port with dust-cap – p. 14
±0.2°C/0.36°F
±0.1°C/0.18°F
Type T −196°C to 400°C/−320.8°F to 752°F
Type K −150°C to 1,370°C/−238°F to 2,498°F
−40°C to 85°C/−40°F to 185°F
Available on 8-channel system.
40°C/
40°F, it is
FOOD TR ACK ER