Power consumption
Standby
• 220 V - 240 V / 50-60 Hz configuration
• 100 V - 120 V / 50-60 Hz configuration
During operation
• 220 V - 240 V / 50-60 Hz configuration
• 100 V - 120 V / 50-60 Hz configuration
Uninterruptible power supply (optional)
UPS Powerware 5115
UPS Powerware 5115
Environmental requirements
Room temperature
Maximum temperature change
Relative humidity
Magnetic field
Sunlight exposure
Transport and storage requirements
Storage
Transport
Warming-up time
Cold start
Warm start
2386B EN 20060926
120 W, 16 A fuse
120 W, 15 A fuse
max. 320 W, 16 A fuse
max. 320 W, 15 A fuse
120 V
ABC ordering code : EGPSE
230 V
ABC ordering code : EGPTG
recommended: 20 °C - 25 °C
allowed: 15 °C - 30 °C
0.5 °C/min.
recommended: 30 % - 60 %
allowed: 15 % - 80 %
compliant with EN 61000-4-8, Level 1
not be operated in full sunlight
IEC 721-3-1 Storage: 1K2, 1M2
MIL-STD 810E Category 1- Basic-Transporta-
tion: method 514.4-1-3.7.4
Vertical, longitudinal- and transversal axis:
frequency range: 5-200 Hz
fully operational after max. 30 min.
fully operational after self-test, on condition
that the CR 30-X:
• has not been switched off for more than
3 minutes.
• has been operational for at least 30 minutes.
CR 30-X
Equipment Information Sheet
49