Errors During Operation
If errors occurs during operation, you can consult the Agfa Service Access
Package (ASAP) messages on the control PC. The ASAP UI is independent
from the NX Software.
Condition
Empty cassette
(no image plate
in cassette)
Error during
read-out of the ID
data
Error during
writing on RF-tag
after scanning
process
Erasure lamp
failed out during
scanning cycle
Image plate
detected in the
digitizer during
self-test
2386B EN 20060926
Control PC message
Cassette error
"Empty cassette – no image
plate detected"
Errors concerning Identification
"Error during read-out of
the ID data. Remove
cassette."
"Error during writing on
RF-tag after succeeded
scanning cycle. Erase the
image plate again."
Digitizer errors
"Failure at Erasure module.
Please reset digitizer. If
error remains, call AGFA-
service technician!
Image Plate must be erased
after repair again."
"Image plate in the digitizer
during self-test detected.
The Image Plate will be
pushed back in the cassette.
Please control if the image
is available on PC and re-
scan the cassette maybe."
Status
indicator
Red, flashing
Remove the cassette
Red, flashing.
Confirm the message
After
with the OK button
unlocking:
and re-identify the
green,
cassette.
constant
Confirm the message
with the OK button on
Red, flashing.
PC, remove the
After
cassette, re-enter the
unlocking:
cassette on PC, remove
green,
the cassette, re-enter
constant
the cassette and re-
erase the cassette.
Restart the digitizer or
Red, flashing
call Agfa service.
After power failure a
remaining image
plate, in the CR 30-X
Red, flashing.
digitizer, will be
detected and the
After
forementioned error
unlocking:
message will be
green,
displayed.
constant
In order to release the
cassette, confirm the
message.
CR 30-X
Action
41